Leybold D40BCS
From Energy Efficient Laboratory Equipment
| Leybold D40BCS | |
| Manfacturer: | Leybold |
| Product Type: | Vacuum pumps |
| Model #: | D40BCS |
| Voltage: | 50Hz: 220-240/380-420 V, 3 Phase
60 Hz: 240-265/415-460 V, 3 Phase |
| Nameplate rated Amps: | 50Hz: 6.9/4 A
60Hz: 6.4/3.7 A |
| Speed: | 47.5 – 51 m3/hr |
Contents |
Energy Efficiency Information
Energy Use and Power
| Power Mode | Definition (temperature, power setting, light features, etc.) | Power Draw VA or W | Can user set mode? | Hours per week (typical)* | kWh per week (typical)* |
| Rated capacity | Nameplate power rating | 1 kW | N/A | N/A | N/A |
| Energy Intensive mode | Power draw under highest setting/energy use | 630-700 W | Yes | 168 | 104.7 kWh |
| Lower power mode 1 | Ultimate Pressure | 630 W | N/A | N/A | N/A |
| Lower power mode 2 | N/A | N/A | N/A | N/A | N/A |
| Stand-by level | Power draw when turned off but connected to power | N/A | N/A | N/A | N/A |
* Typical hours per week and kWh per week are provided for comparison basis only, laboratory values may vary significantly.
Efficiency Features
- N/A
Other Sustainability Features
- N/A
User Comments
Data above measured using EML-2020 Monitor CT electrical meters. Electrical clamps connected to the meter were placed around electrical power wires of the vacuum pumps and their current and voltages were recorded over a span of a week or more. Typically, the Leybold D40BCS was kept at a steady power usage of 625-645 Watts. This pumps was measured in the UCSB ESB Cleanroom. This specific pump used for measurement is an older pump that has been rebuilt multiple times.