To create a new page for a piece of equipment in this category, type in the model name in the input box to the right and click submit. You will be presented with a wiki edit box in which you can fill in values for as many variables that you have information for. More detailed instructions are given on the help page.
shows the change in power consumption from -80'C to -70'C setting.

When using ultra low freezers in laboratories, freezers are often being set at -80'C. However, many samples are safe to be stored in -70'C condition. At the University of Colorado at Boulder (CU-Boulder), the Mark Winey Lab of the Molecular, Cellular, and Developmental Biology (MCDB) Department is storing important stock collections in their ULT freezers set at -70C. Examples in their labs are yeast strains, pelleted cells from tissue culture, ground up yeast cultures, samples to send off for mass spec, ecoli...chemical competent and freezer stocks, purified recombinant protein, pellicles, cell lysates, some antibodies or serum, protein samples.

Contact Us if you want to add more data to media: Chill_Up_Energy_Savings_2010.xls at -70 and -80